Why phosphate matters in plasma-derived therapies
Phosphate is one of the most common buffer components in protein chromatography. Knowing its concentration at a column outlet tells process developers what the column is actually seeing.
Plasma-derived therapies are medicines purified from pooled human plasma. They include immunoglobulins for immune deficiency, albumin and clotting factors. The core separation is cold-ethanol fractionation, developed by Cohn and colleagues in the 1940s. Modern processes then polish the crude proteins on chromatography columns.
Those columns run on carefully controlled buffers, and phosphate is often one of them. This post explains where phosphate measurement can help in process development, and how our phosphate sensing system fits in.
Where phosphate appears in purification
Phosphate can appear at several points in a purification process.
- As a buffer. Ion-exchange steps rely on a set pH and ionic strength, and phosphate buffers are a common way to hold them. They are widely used in immunoglobulin (IgG) purification.
- As the eluent. In hydroxyapatite chromatography, proteins bind to the resin's calcium and phosphate sites. They are released by raising the phosphate concentration, so phosphate is the variable that drives the separation.
- As something to remove. Phosphate used upstream generally has to be washed out before the final formulation step.
Most chromatography systems already measure UV absorbance, pH and conductivity. Conductivity responds to every ion present, so it cannot tell phosphate apart from sodium chloride or other salts. A phosphate-selective reading adds that missing piece.
Our phosphate sensing system
The system has four parts that work together: a sensor, the electronics that read it, solutions to calibrate it, and a solution to rinse and store it.
| Part | What it does | Key facts |
|---|---|---|
| Phosphate sensor | Screen-printed electrochemical sensor for phosphate (H₂PO₄⁻ / HPO₄²⁻ / PO₄³⁻) | Carbon working electrode with an analyte-selective coating; Ag/AgCl reference and counter electrodes; alumina substrate, 24.45 × 7.05 mm. Suits disposable use or continuous measurement. Packs of 10 or 50. |
| Biosensor Measurement Kit | Reads the sensor and sends data to a computer | Single-purpose biosensor board, adapter and FTDI-USB cable. The board detaches from the adapter, so other sensor boards can share it. |
| Phosphate calibration solution kit | Calibrates and checks the sensor | 0.1 mM, 1.0 mM and 10 mM phosphate, 10 ml of each. |
| Sensor rinse solution | Rinses sensors between measurements and before storage | 100 ml or 500 ml bottles. |
For best results, the datasheet recommends using the sensor with ZP potentiostats and our cloud data platform, djuli. Sensors should be stored at 2–15 °C, away from UV light, and have a recommended shelf life of one year.
Where it fits in process development
The system is built for research and development, so its natural place is the process development lab rather than the production floor. There, it can help teams understand and tune their purification steps.
Three uses match what phosphate does in a process and the system's calibrated range of 0.1–10 mM:
- Confirming equilibration. A column is flushed until the outlet matches the inlet buffer. A phosphate reading shows when that has happened, rather than relying on a fixed volume.
- Following buffer changes. When a method switches buffers or runs a desalting step, the outlet reading shows when the old phosphate has cleared.
- Checking wash-out. After phosphate-containing steps, the reading can confirm that phosphate has fallen to a low level before the next step.
A typical bench set-up follows four steps:
- Calibrate the sensor with the 0.1, 1.0 and 10 mM solutions.
- Place the sensor at the column outlet, or test fractions collected from it.
- Record readings through the run with the measurement kit.
- Rinse the sensor with rinse solution, then store it or replace it.
Some steps, such as hydroxyapatite elution, can use phosphate concentrations well above 10 mM. Those applications need a discussion with our team about range, dilution or a custom sensor.
Getting good measurements
A few points make the difference between a useful reading and a misleading one.
- Watch the pH. Phosphate changes form with pH, with pKa values near 2.1, 7.2 and 12.3. Near pH 7, where many chromatography steps run, a small pH shift changes the balance between H₂PO₄⁻ and HPO₄²⁻. Log pH alongside phosphate, and calibrate at a pH close to your process.
- Calibrate regularly. Use the three-point calibration kit before each run, and check against a standard if a run is long.
- Expect protein contact. Protein-rich eluates can coat a sensor surface over time. Rinse between runs, and use the disposable format where carry-over is a concern.
- Use aqueous samples. The sensor is designed for aqueous systems.
Getting started
All four parts are available from the ZP webstore. If your process needs a different concentration range, electrode layout or format, we also make custom sensors, including microwell and microfluidic formats.
Tell us about your application through the contact form, and we will help you choose the right set-up.
Our phosphate sensors are supplied for research and development use.
References
- Cohn EJ et al. Preparation and properties of serum and plasma proteins. J Am Chem Soc 1946;68:459–475.
- Burnouf T. Modern plasma fractionation. Transfus Med Rev 2007;21(2):101–117.
- Cummings LJ, Snyder MA, Brisack K. Protein chromatography on hydroxyapatite columns. Methods Enzymol 2009;463:387–404.
- Zimmer & Peacock. Phosphate sensor technical datasheet, ZPS PO4-003-00078 v.0.10.